A reply
Published in Anaesthesia • Jul 1, 2000
NobleIDNI9P15W65R05S65
Authors:
R. Le Poidevin
Abstract
Intersurgical was notified of this fault in June 1999. The hole in the connector at the end of the oxygen tubing is formed by a pin when the component is produced by the process of injection moulding. The occlusion in this case was caused by a pin in the mould tool becoming damaged, which allowed th...