NobleBlocks

Research indexThesis

Capacitance-based microvolume liquid-level sensor array

Seliskar

eScholarship@McGill (McGill)1 Jan 2006Open access

NobleIDark:/48914/w1/20261105/D9537328DOI10.82308/33594

Abstract

A prototype sensor array was developed for use with laboratory automation to permit closed-loop control of liquid-levels in a multiwell microplate geometry. A simple electrical model for non-contact capacitance-based fluid sensors was extended to describe a fluid-level dependency. The new model shows that a charge-transfer based capacitance transducer employing a liquid-specific calibration can be used to obtain an output signal that varies linearly with the liquid-level when fringe-field effects are negligible. The calibration also compensates for liquid-to-liquid conductivity and permittivity differences. The sensor was tested using sodium chloride (NaCl) and ethanol solutions to simulate the range of conductivity and permittivity typical in biological and chemical research. Measured capacitance was a second-order function of liquid volume due to fringe-field effects and was compensated for by adding a hardware-based calibration. Liquid-volume measurement error averaged 0.2% of the 120mul fill volume with a standard deviation of 0.6% (< mul). The maximum absolute error for all liquids was 2.7% (3mul).

Discussion

More

Suggest a correction or report